System information

Blastware Advanced Module
____________________________________________________
6–16 Blastware Operator Manual
NOTE: The units are set to Pa if the Report Options are Metric, or psi if the Report Options are
set to Imperial.
1. Select the Normal (+/- 100Pa) or Sensitive (+/- 25 Pa) range on Channel 1 and/or 4.
NOTE: Make sure the proper range is selected as the unit cannot dynamically change this while it is
monitoring. After selecting the range the sensitivity will automatically be filled in.
2. Select Trigger Level. Enable or disable any channel including the sound channel as a trigger
source. Assuming it is set as a trigger source the Trigger Level is the level at which the unit
would trigger and start recording data.
Selecting the Trigger Level menu will also allow you to change the recording range between
Sensitive and Normal on the monitor.
3. Edit the values, as required, for the Record Mode, Sample Rate and Record Time. To ensure
the best possible reproduction of the recorded event, a 65536 sps is recommended. Lower
sample rates will trade off between the resolutions and longer record times.
4. If only Channel 1 is to be monitored, turning off all other channels will also allow for longer
recording times.
5. Send the Setup File to the unit.