System information

Blastware Advanced Module
____________________________________________________
6–12 Blastware Operator Manual
Uniaxial 4.5 Hz Geophone (Vertical –
714A4601, Horizontal – 714A4602)
0.16113 V/in/s 0.006344 V/mm/s
Bore Hole Transducer (714A3301) 0.16113 V/in/s 0.006344V/mm/s
Hydrophone Sensor (714A0801) 0.034 V/psi 0.0049 V/kPa
Hydrophone with 100 psi adapter
(714A8201)
0.017 V/psi 0.00244 V/kPa
Series IV
Instantel Sensor
Units
Imperial Metric
Series IV L Microphone (720A1801) 44.4496552 V/psi
(0.0444496 v/mpsi)
0.0064452 V/Pa
Series IV A and C Weight Microphone
(720A5101)
26.2 or 827.40 V/psi 0.0038 or 0.12 V/pa
Series IV Standard Triaxial Geophone
(ISEE, 2-250 Hz) (720A2001)
0.3222605 V/in/s 0.0126774 V/mm/s
Series IV Standard Triaxial Geophone
(DIN, 1-315 Hz) (720A2002)
0.3222605 V/in/s 0.0126774 V/mm/s
Series IV Standard Triaxial Geophone
(DIN, 1-80 Hz) (720A2003)
0.3222605 V/in/s 0.0126774 V/mm/s
Series IV HF Triaxial Geophone
(28 Hz-1 KHz Hz) (720A4701)
0.032226 V/in/s 0.00126774 V/mm/s
Series IV HP Microphone (720A4801)
0.064451998 V/psi
0.00.09340870 kPa
Series IV Accelerometers, .5 g
(720A4901)
Series IV Accelerometers, 50 g
(720A2401)
Series IV Accelerometers, 500 g
(720A0801)
6.4451998 V/g
0.064451998 V/g
0.0064452 V/g
0.6576735 V/m/s2
0.006576735 V/m/s2
0.000657673 V/m/s2
Hydrophone (720A0801) 0.06856596 V/psi 0.00994630 V/kPa
v Gain
Choose the required range and resolution. Click on either button.
vi +/– Trigger
Enter a trigger level if using single shot or continuous record modes.
vii Level
Enter the trigger level.
viii Enable
These checkboxes enable the trigger on a particular channel. You can enable the triggers on
any combination of channels. Recording begins when Event activity exceeds any single set
trigger level.
ix +/– Resolution
The Blastware software automatically calculates the resolution based on the entered sensitivity and
chosen gain.
x +/– Range
The Blastware software automatically calculates the range based on the entered sensitivity and
chosen gain. Refer to this value when setting a trigger level.