User`s manual
2-17
a. To change the default condition, press the
OPTION
key. “NO”
will change to “YES” on the display and you will be able to
select a different tube type when you start a run (see Chapter 6
for more information).
b. Press
ENTER
. The following display appears:
This display lets you change the sample tube time constant.
There are two default values for sample tube time constants on
this display: one for sample tubes with a reaction volume of
100
µ
L, and one for sample tubes with a reaction volume of 50
µ
L. There is a time constant default value for sample tubes
with a reaction volume of 20
µ
L that appears on a separate
display. Press
STEP
to view this display:
c. To change any of these time constant values, press the
appropriate numeric keys, then press
ENTER
. To move the
cursor from one time constant value to the next, or to move
from the first display to the second, press
STEP
.
Press
BACK
to move to the previous display. Or, press
STOP
to return to the first configuration display. From here, you can
return to the main menu by pressing
STOP
twice.
Rxn vol=100uL T= 9.5
Rxn vol= 50uL T= 7.0
Rxn vol= 20uL T= 5.0