Configuring the GeneAmp PCR System 9600
The final procedure in installing the GeneAmp PCR System 9600
is checking the instrument configuration. The configuration
parameters include:
time and date
run-time printer mode (on or off)
run-time beeper (on or off)
pause time-out limit
allowed setpoint error
idle state setpoint
temperature interval that triggers the clock start
sample tube time constants
Access the configuration parameters as follows:
1
The Utilities menu (shown below) should be displayed. If not,
access the main menu, and select UTIL.
2
Press
OPTION
to move the cursor to CONFIG, then press
ENTER
.
The first configuration display is shown. Refer to the following
pages for instructions on editing configuration parameters. For
more detailed instructions, see Chapter 7, Utilities..
Select function
D
IR-CONFIG-DIAG-DEL