User`s manual
2-7
4
Connect one end of the cable (Applied Biosystems Part No.
N801-0421) to the 25-pin (D shell) RS-232C connector on the side
panel of the GeneAmp PCR System 9600. Connect the other end
of the cable to the connector on the serial adapter card you just
installed.
Note
When you are printing, the printer ON LINE/OFF LINE button
must be in the ON LINE position. See page 3-3 of the Epson LX-810
User's manual.
When setting the DIP switches on the serial interface cards,
remember that the “ON” setting is the same as the “CLOSED”
setting, and the “OFF” setting is the same as the “OPEN” setting.
The correct DIP switch settings on the Epson C82302 serial
adapter card are:
SW1 SW2
1: OFF 1: ON
2: OFF 2: ON
3: OFF 3: OFF
4: OFF 4: OFF
5: OFF 5: OFF
6: ON 6: OFF
7: OFF 7: OFF
8: OFF 8: OFF