User`s manual

9-4
Upon power-up or
returning to an
instrument that has
been running
unattended, the
following display
appears:
This is normal if the power
has been interrupted or
shut off while running a
program.
Review the history file to
determine when the
power failed. Note that the
system comes up in a 4-
degree soak to save any
samples. The display
shows how long power
was interrupted.
The printer is not
printing run-time data.
The run-time printer mode
configuration parameter is
not turned on.
Turn the run-time printer
mode on.
The printer has been in-
correctly set up.
Check the printer set-up.
Bad Printer cable. Replace the printer cable.
When you turn on the
instrument, the
following display
appears:
Possible controller failure. Contact an Applied
Biosystems Service
Engineer.
Table 9-1 Troubleshooting Information
(continued)
Problem Possible Cause Check and / or Remedy
Power Fail x min
Review history? YES
Call Service