Changing External Fuses
There are two slow-blow fuses accessible from the left side of the
instrument. Use the following procedure to change these fuses.
Use 15-Amp slow-blow fuses for the 100 and 120 VAC
instruments. Use 8-Amp slow-blow fuses for the 220-240 VAC
instruments. See Chapter 10 for information on power
requirements.
1
Turn off the GeneAmp PCR System 9600 and disconnect the power
cord from the instrument.
2
The fuses are located next to the instrument power cord connection
(see Figure 8-8).
Figure 8-8
Removing the fuses.
3
Remove the fuse(s) with a screwdriver. Replace the fuse(s) with the
same type and rating of fuse.
4
Replace the instrument power cord.
Fuse Holders
Instrument Power Cord Connection
F1 F2