User`s manual
7-44
Figure 7-10
A data record display of a CYCL run.
The above example tells you that this data record is record #6 of
CYCL program #2. There is no method number because this
program was not linked in a method. This record also tells you
that this is setpoint #2 of cycle #1, the ending setpoint
temperature was 25.2 degrees C, and the elapsed time was ten
seconds. If this were a data record for a HOLD program, there
would be no cycle number or setpoint number. If an error had
occurred at this point in the run, an error message would also
have appeared on this display.
Displaying the History file
To select Diagnostic Test #1 and display the History file:
1
Select DIAG on the Utilities menu as described on page 7-3.
The following display appears:
2
Press
ENTER
to select Diagnostic Test #1.
The following display appears:
Enter Diag Test #1
REVIEW HISTORY FILE
History 7 recs
A
LL-STAT-ERROR-PRINT
Program
Type
Program #
Record #
Ending Setpoint
Temperature
Cycle # Setpoint #
Elapsed
Time
C# 2 25.2C 6
Cyc 1 Setpt 2 0:10