User`s manual

7-34
CAUTION
If your instrument has software version 1.04 or lower,
configure the THIN tube time constant for a default response time
of 12.0 seconds for 20-µL reaction volume. When initiating a PCR
program, specify a reaction volume of 20 µL for all experiments
using Thin-Walled GeneAmp Reaction Tubes.
If your instrument has software version 1.05 or higher, the
response time of 12.0 seconds for all three reaction volumes is
correct for the Thin-Walled GeneAmp Reaction Tubes.
Printing the Configuration Parameters
The PRINT selection of the Configuration function lets you print
all of the configuration parameters. To print the parameters:
1
Select CONFIG on the Utilities menu as described on page 7-3.
The following display appears:
2
Press
OPTION
to move the cursor to PRINT, then press
ENTER
.
As the configuration parameters are printed, the following display
appears:
3
When printing has been completed, the first Configuration display
appears. Press the
STOP
key to return to the Main menu.
Configuration
E
DIT-PRINT
Configuration
printing...