User`s manual

7-30
Setting Sample Tube Time Constants
There are two additional parameters in the Configuration function
that let you change the time constants for MicroAmp Reaction
Tubes and Thin Walled GeneAmp Reaction Tubes. We
recommend that you do not change the time constant values
unless instructed to do so by future documentation.
MICRO Time Constant
To change the sample tube time constant value for MicroAmp
Reaction Tubes:
1
Select CONFIG on the Utilities menu as described on page 7-3.
The following display appears:
2
Press
ENTER
, then press
STEP
until the following display appears:
3
The default condition of this parameter is no. Press the
OPTION
key. NO on the display will change to DEFAULTS.
4
Press the
OPTION
key again. DEFAULTS will change to
MODIFY. Pressing the
OPTION
key again will change MODIFY
back to NO.
Configuration
E
DIT-PRINT
Set MICRO tube time
constants? NO