User`s manual

7-27
2
Press
ENTER
, then press
STEP
five times.
The following display appears:
This example shows the default temperature interval of 1 degree C.
This means that the setpoint clock will start counting within 1
degree of the setpoint temperature.
3
To enter a temperature interval different from the default value,
press the appropriate numeric keys, then press
ENTER
.
The range of valid entries for the temperature interval is 0.5 to 1.0
degrees C. If you decrease the temperature interval from 1.0
degree C, the clock will start closer to the setpoint temperature.
Note
To maintain portability of PCR programs, we highly recommend
that you use the default value of 1.0 degree C. The hold time decreases
as the temperature interval for the setpoint clock decreases. Change
this parameter only for specific experiments.
4
Press
STEP
to display the next configuration parameter or press
BACK
to display the previous configuration parameter. Press
STOP
to return to the first Configuration display.
Start clock within
1.0
C of setpoint