User`s manual
6-12
Running a Stored Program
To run a program that you have previously stored:
1
Access the main menu.
CAUTION
Before you start a run, make sure the “Start clock
within x.xC of setpoint” parameter in the Configuration function is
set correctly. To do this, select “UTIL” on the main menu, then
select “CONFIG” on the Utilities menu. Press ENTER, then press
the STEP key five times to view the “Start clock...” parameter. See
Chapter 7 for complete instructions.
2
With the cursor positioned under RUN, press
ENTER.
The following display appears:
3
Select the desired program by pressing the appropriate numeric
keys, then
ENTER
.
The following display appears:
This display lets you select the type of sample tube you are using
and enter the reaction volume for that tube. The default tube type is
the MicroAmp Reaction Tube, as indicated on the display by
“MICRO.” The default reaction volume is 100
µ
L.
Select Option 9600
R
UN-CREATE-EDIT-UTIL
Run
Enter program #
Select tube MICRO
Reaction vol? 100
uL