User`s manual

5-43
These two displays are used to set the temperature
and hold time for the third setpoint (Setpt #3) and
change the time and temperature values in
successive cycles, if desired.
To change these displays, use the same procedure
as described for Setpoint #1.
A display will appear for each setpoint up until the
total number of setpoints specified in the first
program display. For each setpoint, you will be
asked whether you want to change the time or
temperature values in successive cycles.
Specifies the number of times the complete cycle
will be repeated. Type the number of cycles (1-99),
then press ENTER.
This display also allows you to program a pause
after a setpoint and cycle that you will specify in
subsequent displays. A pause would be required,
for example, to add a reagent. A pause may also be
used to add or remove sample tubes during a run.
Press the OPTION key to change the NO response
(default) to a YES and vice versa. When the desired
response is displayed, press ENTER.
Setpt #3 72.0
C
Hold for 0:30
Setpt #3 72.0
C
Change time/temp? NO
Setpt #n 0.0C
Hold for 0:30
Total cycles = 25
Pause during run? NO