User`s manual

31
Understanding Your Test Results
The EasyPlus mini R13N Blood Glucose test strips are plasma referenced and calibrated for easier
comparison to lab results. The normal fasting blood glucose range for an adult with diabetes is 70-130 mg/dL
(3.9~7.2 mmol/L)*.
Two hours after meals, the blood glucose range for an adult with diabetes is less than 180
mg/dL (10.0 mmol/L).
For further queries about diabetes: please consult your doctor for the blood glucose
range appropriate for you.
Reference: American Diabetes Association. Standards of medical care in diabetes-Table 10. Diabetes care. 2011; Vol. 34, Suppl 1, S21.
Unusual Test Results
If your test result doesn’t match the way you feel, please follow these steps:
1. Run a control solution test, Chapter 2 "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3 "Testing Your Blood glucose."
3. If your test results still don’t reflect the way you feel, call your healthcare professional immediately.
1. Extremely high humidity may affect the test results. A relative humidity greater than
90% may cause inaccurate results.
2. Hematocrit below 30% may cause higher results. Hematocrit above 55% may cause
lower results.
3. Some studies have shown that electromagnetic fields may affect results. Do not test
near an operating microwave oven.