User`s manual

45
Limitations
The test strips are used for fresh capillary whole blood samples.
1. DO NOT use neonate blood sample.
2. Extreme humidity may affect the results. A relative humidity greater than 90 % may cause incorrect results.
3. The system should be used at temperatures between 50°F and 104°F (10°C and 40°C). Outside this range,
the system may get incorrect results.
4. DO NOT reuse the test strips. The test strips are for single use only.
5. Hematocrit: The hematocrit between 20% and 60% will not affect the results. Hematocrit below 20% may
cause higher results. Hematocrit above 60% may cause lower results.
6. Altitude up to 3,048 meters above sea level has no effect on readings.
Healthcare Professionals – Please note these additional Limitations
1. Patients undergoing oxygen therapy may have inaccurate results.
2. If the patient has the following conditions, the result may fails:
Severe dehydration
Severe hypotension (low blood pressure)
Shock
A state of hyperglycemic-hyperosmolar state (with or without ketosis)
3. Lipemic samples: Cholesterol level up to 500 mg/dL (12.92 mmol/L) and triglycerides up to 3,000 mg/dL
(33.6mmol/L) do not affect the results. Grossly lipemic patient samples have not been tested and are not
recommended for testing with EASYMAX
®
MU Glucose Meter.
4. Critically ill patients should not be tested with EASYMAX
®
MU glucose meters.
5. Interfering Substances depend on the concentration. The below substances up to the test concentration
will not affect the test results.