Understanding Your Test Results
The EASYMAX
®
MU Blood Glucose test strips are whole-blood referenced and calibrated for easier
comparison to lab results. The American Diabetes Association recommends a post-meal glucose level of less
than 180 mg/dL (10 mmol/L) and a pre-meal glucose of 70–130 mg/dL (3.88-7.22 mmol/L)*.
Unusual Test Results
If your test result does not match the way you feel, please follow these steps:
1. Run a control test, Chapter 2, "Control Solution Testing."
2. Repeat a blood glucose test, Chapter 3, "Testing Your Blood glucose."
3. If your test results still do not reflect the way you feel, call your doctor immediately.
Reference: American Diabetes Association. Standards of medical care in diabetes-Table 10. Diabetes
care. 2011; Vol. 34, Suppl. 1, S21.
1. Extremely high humidity may affect the test results.
A
relative humidity greater than 90%
may cause inaccurate results.
2. Hematocrit below 20% may cause higher results. Hematocrit above 60% may cause
lower results.
3. Some studies have shown that electromagnetic fields may affect results. Do not test
near an operating microwave oven.