User`s manual

17
Chapter 2: Control Solution Testing
Why Run a Control Solution Test
We recommend that you run the EASYMAX
®
MU control test because it lets you know that your meter and
test strips are working properly to give reliable results. You should run the control solution tests when:
You use the EASYMAX
®
MU Blood Glucose Meter for the first time.
You open a new bottle of test strips.
You think the meter or test strips may be working incorrectly.
You drop the meter.
You have repeated a test and the test results are still lower or higher than expected.
You are practicing the test procedure.
Professional users are instructed to follow federal, state, and local guidelines.