User's Manual
Table Of Contents
- The EasyMax® S1 Blood Glucose Meter
- Inserting Batteries
- Setting The Year and Hour Clock
- Why Set High Range Limit
- Control solutions can stain clothing. If you spill it, wash your clothes with soap and water.
- Close the bottle tightly after every use.
- Left over control solution should not be added back into the control bottle.
- Store control solution at room temperature, between 36 F - 86 F (2 C~30 C). Do not freeze.
- If you would like to purchase EASYMAX® S1 level 2 and level 3 Control Solutions, please contact your local dealer.
- The last 40 test results can be transferred to your phone with NFC. To Transfer data, open GlucoManager™ app as well as the meter can be turned on or off.
- Super Sani-Cloth germicidal wipe contains active ingredients: n-Alkyl (60% C14, 30%C16, 5%C12, 5%C18) dimethyl benzyl ammonium chlorides and n-Alkyl (68%C12, 32%C14) dimethyl ethylbenzyl ammonium chlorides and they have been shown to be safe for use w...
- 4 year product life is for properly cleaning and disinfection. After 4 years, the meter must be replaced with a new meter.
- Replace the batteries first.
- System error
- Refer to pages 15 and 40. If this error message appears again, please contact your local dealer.
- There may be a problem with the meter.
- Repeat the test with a new strip.
- Volume detector error
- If Err appears again, please contact your local dealer.
- The volume of blood or control solution is NOT enough.
- Repeat the test with a new strip.
- Humidified / Used strips
- Refer to pages 28-29 for information on sample application.
- The meter has detected a problem with the test strip.
- Replace the batteries first.
- Memory Error
- If ERROR 005 appears again, please contact your local dealer.
- Re-check your glucose level.
- Test result is higher than 630 mg/dL (35 mmol/L).
- If the result is HI again, obtain and follow instructions from your healthcare professional without delay.
- This may require immediate treatment according to your healthcare professional’s recommendations. Although this message could be due to a test error, it is safer to treat first and then do another test.
- The test result is lower than 20 mg/dL (1.1 mmol/L).
- The “Ht” and thermometer icon appears. Temperature is too high, outside the required range of 5°C - 45°C (41°F - 113°F). This alerts users that an incorrect result may occur if the test continues.
- Relocate the meter to a location with temperature between 5°C - 45°C (41°F - 113°F).
- Relocate the meter to a location with temperature between 5°C - 45°C (41°F - 113°F).
- The “Lt” and thermometer icon appears. Temperature is too low, outside the required range of 5°C - 45°C (41°F - 113°F). This alerts users that an incorrect result may occur if the test continues.
- You can still perform a blood glucose test and get an accurate test result.
- No result in memory
- The meter doesn’t run any test yet.
- Replace the new batteries.
- Low power
- The meter batteries do not have enough power to perform a test.
- Repeat the upload.
- System error
- There may be a problem with data transfer.
- Refer to pages 41. If this error message appears again, please contact your local dealer.
- EASYMAX® S1 SMBG System,
- EASYMAX® S1 Blood Glucose Test Strips,
- EASYMAX® S1 Blood Glucose Meter,
- EASYMAX® S1 Level 2 Control Solution,
- EASYMAX® S1 Level 3 Control Solution
- Reference:
- * American Diabetes Association. Standards of medical care in diabetes-Table 10 Diabetes care. 2011; Vol. 34, Suppl. 1, S21.
- Manufacturer:
- EPS BIO TECHNOLOGY CORP.
- TEL:+886-3-6686868 FAX:+886-3-6686866
- E-mail:info@epsbio.com.tw Website:http://www.epsbio.com
- Lancing Device
- Meets the requirements of MDD 93/42/EEC
- Lancet
- Meets the requirements of MDD 93/42/EEC
- SAE HAN MEDICAL CORP.
- TEL:82-31-923-4330 FAX:82-31-923-4331 E-mail:saehan@saehanmed.com
- P/N: 71800580A_0129A_01
56
Limitations
The test strips are used for fresh venous blood and capillary whole blood samples.
1. DO NOT use neonate blood sample.
2. Not to be used for diagnosis or screening of diabetes.
3. Alternate site testing with this system can be used only during steady-state blood glucose conditions.
4. Measurements from alternate site testing should never be used to calibrate a continuous glucose monitor
(CGM) or entered into insulin dose calculators for insulin dosing recommendations.
5. Extreme humidity may affect the results. A relative humidity greater than 90 % may cause incorrect results.
6. The system should be used at a temperature between 50°F and 104°F (10°C and 40°C). Outside this
range, the system may get incorrect results.
7. DO NOT reuse the test strips. The test strips are for single use only.
8. Hematocrit: The hematocrits between 20% and 60% will not significantly affect the results. Hematocrit
below 20% may cause higher results. Hematocrit above 60% may cause lower results. If you do not know
your hematocrit level, please consult with your healthcare professional.
9. Altitude up to 10,000 feet above sea level has no effect on readings.
10. The below substances up to the test concentration will not affect the test results.
Interfering substance Concentration tested up to (mg/dL)
Gentisic Acid
6
Ascorbic Acid
4
Ibuprophen
50
Methyldopa
2
Sodium Salicylate
50
Tetracycline
1.5