User's Manual

Table Of Contents
52
Message
What it means?
What to do?
The Htand thermometer icon
appears. Temperature is too high,
outside the required range of 5°C -
45°C (41°F - 113°F). This alerts
users that an incorrect result may
occur if the test continues.
Relocate the meter to a location with
temperature between 5°C - 45°C (41°F
- 113°F).
The Ltand thermometer icon
appears. Temperature is too low,
outside the required range of 5°C -
45°C (41°F - 113°F). This alerts
users that an incorrect result may
occur if the test continues.
Relocate the meter to a location with
temperature between 5°C - 45°C (41°F
- 113°F).