Cleaning and Disinfecting Your Meter and Lancing Device
Choosing the disinfectant
The recommended EPA-registered disinfected product is as follow:
PDI® Super SANI-CLOTH® Germicidal disposable wipe (EPA Reg. No.:9480-4)
Super Sani-Cloth germicidal wipe contains active ingredients: n-Alkyl (60% C14, 30%C16, 5%C12, 5%C18)
dimethyl benzyl ammonium chlorides and n-Alkyl (68%C
12, 32%C14) dimethyl ethylbenzyl ammonium
chlorides and they have been shown to be safe for use with the EASYMAX
®
S1 Blood Glucose Meter, but any
other disinfectant product with the EPA registration number may be used on this device.
Please purchase in retail stores like Walmart and Office Depot. You could also purchase on the PDI
website:http://www.pdipdi.com/
or online retail sites like amazon.com, medline.com, and Expression medical
supply Inc. http://www.exmed.net/.
Cleaning and Disinfection Instruction
Please keep the meter and lancing device free of dirt, dust, bloodstain, and water stains. Please follow the
following guidelines to clean and disinfect your meter and lancing device.
The meter and lancing device have been validated so that they can withstand cleaning for up to 9 times per
day.
After every use, follow the cleaning instruction to prevent the growth of any microorganism and also to
help improving the effectiveness of disinfection. Then follow disinfection instruction once per week which
could effectively kill blood borne pathogens, such as viral hepatitis and prevent cross-contamination. If the
devices are being operated by a second person who is providing testing assistance to the user, the meter and
lancing device should be disinfected prior to use by the second person.