- The EasyMax® S1 Blood Glucose Meter
- Inserting Batteries
- Setting The Year and Hour Clock
- Why Set High Range Limit
- Control solutions can stain clothing. If you spill it, wash your clothes with soap and water.
- Close the bottle tightly after every use.
- Left over control solution should not be added back into the control bottle.
- Store control solution at room temperature, between 36 F - 86 F (2 C~30 C). Do not freeze.
- If you would like to purchase EASYMAX® S1 level 2 and level 3 Control Solutions, please contact your local dealer.
- The last 40 test results can be transferred to your phone with NFC. To Transfer data, open GlucoManager™ app as well as the meter can be turned on or off.
- Super Sani-Cloth germicidal wipe contains active ingredients: n-Alkyl (60% C14, 30%C16, 5%C12, 5%C18) dimethyl benzyl ammonium chlorides and n-Alkyl (68%C12, 32%C14) dimethyl ethylbenzyl ammonium chlorides and they have been shown to be safe for use w...
- 4 year product life is for properly cleaning and disinfection. After 4 years, the meter must be replaced with a new meter.
- Replace the batteries first.
- System error
- Refer to pages 15 and 40. If this error message appears again, please contact your local dealer.
- There may be a problem with the meter.
- Repeat the test with a new strip.
- Volume detector error
- If Err appears again, please contact your local dealer.
- The volume of blood or control solution is NOT enough.
- Repeat the test with a new strip.
- Humidified / Used strips
- Refer to pages 28-29 for information on sample application.
- The meter has detected a problem with the test strip.
- Replace the batteries first.
- Memory Error
- If ERROR 005 appears again, please contact your local dealer.
- Re-check your glucose level.
- Test result is higher than 630 mg/dL (35 mmol/L).
- If the result is HI again, obtain and follow instructions from your healthcare professional without delay.
- This may require immediate treatment according to your healthcare professional’s recommendations. Although this message could be due to a test error, it is safer to treat first and then do another test.
- The test result is lower than 20 mg/dL (1.1 mmol/L).
- The “Ht” and thermometer icon appears. Temperature is too high, outside the required range of 5°C - 45°C (41°F - 113°F). This alerts users that an incorrect result may occur if the test continues.
- Relocate the meter to a location with temperature between 5°C - 45°C (41°F - 113°F).
- Relocate the meter to a location with temperature between 5°C - 45°C (41°F - 113°F).
- The “Lt” and thermometer icon appears. Temperature is too low, outside the required range of 5°C - 45°C (41°F - 113°F). This alerts users that an incorrect result may occur if the test continues.
- You can still perform a blood glucose test and get an accurate test result.
- No result in memory
- The meter doesn’t run any test yet.
- Replace the new batteries.
- Low power
- The meter batteries do not have enough power to perform a test.
- Repeat the upload.
- System error
- There may be a problem with data transfer.
- Refer to pages 41. If this error message appears again, please contact your local dealer.
- EASYMAX® S1 SMBG System,
- EASYMAX® S1 Blood Glucose Test Strips,
- EASYMAX® S1 Blood Glucose Meter,
- EASYMAX® S1 Level 2 Control Solution,
- EASYMAX® S1 Level 3 Control Solution
- Reference:
- * American Diabetes Association. Standards of medical care in diabetes-Table 10 Diabetes care. 2011; Vol. 34, Suppl. 1, S21.
- Manufacturer:
- EPS BIO TECHNOLOGY CORP.
- TEL:+886-3-6686868 FAX:+886-3-6686866
- E-mail:info@epsbio.com.tw Website:http://www.epsbio.com
- Lancing Device
- Meets the requirements of MDD 93/42/EEC
- Lancet
- Meets the requirements of MDD 93/42/EEC
- SAE HAN MEDICAL CORP.
- TEL:82-31-923-4330 FAX:82-31-923-4331 E-mail:saehan@saehanmed.com
- P/N: 71800580A_0129A_01
30
Alternate Site Testing (AST)
Understanding Alternate Site Testing
Consult your health care professional before you begin using the palm or forearm for testing. Blood glucose
test results oS1ained from your palm or forearm may differ significantly from fingertip samples.
We strongly recommend that you:
Do AST ONLY in the following intervals:
In a pre-meal or fasting state (more than 2 hours since the last meal).
Two hours or more after taking insulin.
Two hours or more after exercise.
Do NOT use AST if:
You think your blood glucose is low.
You are unaware of hypoglycemia.
Your AST results do not match the way you feel.
You are testing for hyperglycemia.
Your routine glucose results are often fluctuating.
If you are pregnant.
What is AST?
Besides the fingertip, you can test your palm or
forearm.
What is the advantage of AST?
You have the option of testing other places on
your body besides the fingertip.