User's Manual
Table Of Contents
- The EasyMax® S1 Blood Glucose Meter
- Inserting Batteries
- Setting The Year and Hour Clock
- Why Set High Range Limit
- Control solutions can stain clothing. If you spill it, wash your clothes with soap and water.
- Close the bottle tightly after every use.
- Left over control solution should not be added back into the control bottle.
- Store control solution at room temperature, between 36 F - 86 F (2 C~30 C). Do not freeze.
- If you would like to purchase EASYMAX® S1 level 2 and level 3 Control Solutions, please contact your local dealer.
- The last 40 test results can be transferred to your phone with NFC. To Transfer data, open GlucoManager™ app as well as the meter can be turned on or off.
- Super Sani-Cloth germicidal wipe contains active ingredients: n-Alkyl (60% C14, 30%C16, 5%C12, 5%C18) dimethyl benzyl ammonium chlorides and n-Alkyl (68%C12, 32%C14) dimethyl ethylbenzyl ammonium chlorides and they have been shown to be safe for use w...
- 4 year product life is for properly cleaning and disinfection. After 4 years, the meter must be replaced with a new meter.
- Replace the batteries first.
- System error
- Refer to pages 15 and 40. If this error message appears again, please contact your local dealer.
- There may be a problem with the meter.
- Repeat the test with a new strip.
- Volume detector error
- If Err appears again, please contact your local dealer.
- The volume of blood or control solution is NOT enough.
- Repeat the test with a new strip.
- Humidified / Used strips
- Refer to pages 28-29 for information on sample application.
- The meter has detected a problem with the test strip.
- Replace the batteries first.
- Memory Error
- If ERROR 005 appears again, please contact your local dealer.
- Re-check your glucose level.
- Test result is higher than 630 mg/dL (35 mmol/L).
- If the result is HI again, obtain and follow instructions from your healthcare professional without delay.
- This may require immediate treatment according to your healthcare professional’s recommendations. Although this message could be due to a test error, it is safer to treat first and then do another test.
- The test result is lower than 20 mg/dL (1.1 mmol/L).
- The “Ht” and thermometer icon appears. Temperature is too high, outside the required range of 5°C - 45°C (41°F - 113°F). This alerts users that an incorrect result may occur if the test continues.
- Relocate the meter to a location with temperature between 5°C - 45°C (41°F - 113°F).
- Relocate the meter to a location with temperature between 5°C - 45°C (41°F - 113°F).
- The “Lt” and thermometer icon appears. Temperature is too low, outside the required range of 5°C - 45°C (41°F - 113°F). This alerts users that an incorrect result may occur if the test continues.
- You can still perform a blood glucose test and get an accurate test result.
- No result in memory
- The meter doesn’t run any test yet.
- Replace the new batteries.
- Low power
- The meter batteries do not have enough power to perform a test.
- Repeat the upload.
- System error
- There may be a problem with data transfer.
- Refer to pages 41. If this error message appears again, please contact your local dealer.
- EASYMAX® S1 SMBG System,
- EASYMAX® S1 Blood Glucose Test Strips,
- EASYMAX® S1 Blood Glucose Meter,
- EASYMAX® S1 Level 2 Control Solution,
- EASYMAX® S1 Level 3 Control Solution
- Reference:
- * American Diabetes Association. Standards of medical care in diabetes-Table 10 Diabetes care. 2011; Vol. 34, Suppl. 1, S21.
- Manufacturer:
- EPS BIO TECHNOLOGY CORP.
- TEL:+886-3-6686868 FAX:+886-3-6686866
- E-mail:info@epsbio.com.tw Website:http://www.epsbio.com
- Lancing Device
- Meets the requirements of MDD 93/42/EEC
- Lancet
- Meets the requirements of MDD 93/42/EEC
- SAE HAN MEDICAL CORP.
- TEL:82-31-923-4330 FAX:82-31-923-4331 E-mail:saehan@saehanmed.com
- P/N: 71800580A_0129A_01
2
Dear EASYMAX
®
S1 SMBG System Owner,
Thank you for using the EASYMAX
®
S1 Self-Monitoring Blood Glucose (SMBG) System. We designed this
system to be dependable, easy-to-use, compact, lightweight and portable to help you monitor your blood
glucose on a regular basis.
Please read this manual thoroughly before you begin testing. This manual provides you and your diabetes
care team with important information and step-by-step direction to use the EASYMAX
®
S1 Self-Monitoring
Blood Glucose System.
Thanks again for choosing the EASYMAX
®
S1 SMBG.
Intended Use
The EASYMAX
®
S1 Self Monitoring Blood Glucose Test System is intended for the quantitative
measurement of glucose in fresh capillary whole blood, venous blood and neonate blood samples drawn from
the fingertip, palm, or forearm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use
at home (over the counter [OTC]) by a single patient with diabetes and should not be shared, as an aid to
monitor the effectiveness of diabetes control. The system is not to be used on neonates, nor for the diagnosis
of, or screening for diabetes mellitus. Alternate site testing can be only used during steady-state blood glucose
conditions.