User's Manual
Table Of Contents
- The EASYMAX® Tag Blood Glucose Meter
- Inserting A Battery
- Setting The Year and Hour Clock
- Using EASYMAX® Tag Blood Glucose Test Strips
- Control solutions can stain clothing. If you spill it, wash your clothes with soap and water.
- Close the bottle tightly after every use.
- Left over control solution should not be added back into the control bottle.
- Store control solution at room temperature, between 36 F - 86 F (2 C~30 C). Do not freeze.
- If you would like to purchase EASYMAX® Tag level 2 and level 3 Control Solutions, please contact your local dealer.
- Viewing & Deleting Test Results
- Running With Your Smartphone
- Test results can be transferred to your mobile with NFC. To Transfer data, the meter can be turned on or off.
- Super Sani-Cloth germicidal wipe contains active ingredients: n-Alkyl (60% C14, 30%C16, 5%C12, 5%C18) dimethyl benzyl ammonium chlorides and n-Alkyl (68%C12, 32%C14) dimethyl ethylbenzyl ammonium chlorides and they have been shown to be safe for use w...
- 4 year product life is for properly cleaning and disinfection. After 4 years, the meter must be replaced with a new meter.
- Replace the batteries first.
- System error
- Refer to pages 15 and 28. If this error message appears again, please contact your local dealer.
- There may be a problem with the meter.
- Repeat the test with a new strip.
- Volume detector error
- If Err appears again, please contact your local dealer.
- The volume of blood or control solution is NOT enough.
- Repeat the test with a new strip.
- Humidified / Used strips
- Refer to pages 27-28 for information on sample application.
- The meter has detected a problem with the test strip.
- Replace the batteries first.
- Memory Error
- If ERROR 005 appears again, please contact your local dealer.
- Re-check your glucose level.
- Test result is higher than 630 mg/dL (35 mmol/L).
- If the result is HI again, obtain and follow instructions from your healthcare professional without delay.
- This may require immediate treatment according to your healthcare professional’s recommendations. Although this message could be due to a test error, it is safer to treat first and then do another test.
- The test result is lower than 20 mg/dL (1.1 mmol/L).
- The “Ht” and thermometer icon appears. Temperature is too high, outside the required range of 10°C - 40°C (50°F - 104°F). This alerts users that an incorrect result may occur if the test continues.
- Relocate the meter to a location with temperature between 10°C - 40°C (50°F - 104°F).
- Relocate the meter to a location with temperature between 10°C - 40°C (50°F - 104°F).
- The “Lt” and thermometer icon appears. Temperature is too low, outside the required range of 10°C - 40°C (50°F - 104°F). This alerts users that an incorrect result may occur if the test continues.
- Replace the new batteries.
- Low power
- The meter batteries do not have enough power to perform a test.
- You can still perform a blood glucose test and get an accurate test result.
- No result in memory
- The meter doesn’t run any test yet.
- EASYMAX® Tag SMBG System,
- EASYMAX® Tag Blood Glucose Test Strips,
- EASYMAX® Tag Blood Glucose Meter,
- EASYMAX® Tag Bluetooth Dongle,
- EASYMAX® Tag Level 2 Control Solution,
- EASYMAX® Tag Level 3 Control Solution
- Reference:
- * American Diabetes Association. Standards of medical care in diabetes-Table 10 Diabetes care. 2011; Vol. 34, Suppl. 1, S21.
- Manufacturer:
- EPS BIO TECHNOLOGY CORP.
- TEL:+886-3-6686868 FAX:+886-3-6686866
- E-mail:info@epsbio.com.tw Website:http://www.epsbio.com
- Lancing Device
- Meets the requirements of MDD 93/42/EEC
- Lancet
- Meets the requirements of MDD 93/42/EEC
- SAE HAN MEDICAL CORP.
- TEL:82-31-923-4330 FAX:82-31-923-4331 E-mail:saehan@saehanmed.com
- P/N: 71800572A_0121A_01
42
Cleaning
Instruction: All blood and other body fluids must be thoroughly cleaned from surfaces and
objects before disinfection by the germicidal wipe. Open, unfold and use first germicidal wipe to remove heavy
soil after every use.
Disinfection Instruction
: Unfold a clean wipe and thoroughly wet all the surface of the meter, including the
strip port and other connection port. Unfold a clean wipe and thoroughly wet all the surface of the lancing
device, including cap or AST cap if used. Treated area must remain visibly wet for a full 2 minutes. Let the
devices air dry for 0.5 minute. Do disinfection once per week.
Do Not:
Get any moisture in the test strip slot.
Spray any cleaning solution directly onto the
meter.
Put the meter under water or liquid.
Pour liquid into the meter.
Do:
Make sure the meter is turned off during cleaning
and disinfection.
Keep the test strip vials tightly c
losed when
performing the cleaning and disinfection
procedures because the fumes from the
disinfectant may affect the performance of the strip
After cleaning or disinfection, please perform the
physical appearance and performance check of
devices.
Please follow the instruction on page 41 to 43.