User guide
Temperature Verification System USB – Single-channel — Operating manual
35
9.3.2 Mastercycler
9.3.3 Thermomixer and ThermoStat
RT = room temperature
9.3.4 ThermoMixer C, ThermoMixer F1.5, ThermoMixer FP and ThermoStat C
Mastercycler family Temperature Error limits
Mastercycler nexus 35 °C ±0,6 K
95 °C ±1,0 K
Mastercycler ep 35 °C ±0,6 K
95 °C ±1,0 K
Mastercycler pro 35 °C ±0,6 K
95 °C ±1,0 K
Temperature Thermomixer
comfort
Thermomixer
compact
ThermoStat plus
0 °C ——— ——— ±1,0 K
10 °C below RT ±2,0 K ——— ———
37 °C ±0,5 K ±1,0 K ±0,5 K
56 °C ±2,0 K ±2,0 K ±1,0 K
75 °C ±2,0 K ±2,0 K ±1,0 K
90 °C ±2,0 K ±2,0 K ±1,0 K
Temperature Error limits
20 °C-45 °C ±1,0 K
<20 °C and >45 °C ±1,0 K