5.1 Verifying the Mastercycler nexus, pro and ep
5.1.1 Overview
The verification program checks the temperature accuracy of the thermoblock.
The Mastercycler heats the thermoblock to 95 °C and then to 35 °C. In the process, the user will
be prompted to measure the temperature in 6 different block positions that are shown on the
display. The verification procedure lasts approx. 30 minutes.
The Mastercycler outputs the measuring results on the display and on a printer or storage
medium.
The CycleManager generates a protocol of the measurement on the monitor. The protocol can be
printed or saved as a text file using the File menu.
The following Mastercyclers save the measuring results on a MultiMediaCard:
Mastercycler ep
Mastercycler pro
The following Mastercyclers save the measuring results on a USB stick:
Mastercycler pro
Mastercycler nexus
The Mastercycler must be adjusted if the measured values are outside of the maximum
permissible error.
The maximum permissible errors for the Temperature Verification System USB – Single-channel
and the devices are listed in the technical data.
5.1.2 Verifying the Mastercycler nexus and Mastercycler pro
Prerequisites
The Mastercycler is switched on.
The Measuring device with the temperature sensor is connected to the cycler or the control
panel and switched on.
Hint!
Only users who are logged in as User or Administrator can verify the
Mastercycler nexus, pro or ep.
Hint!
The Mastercycler nexus flat cannot be adjusted using the Temperature
Verification System USB – Single-channel.
Hint!
All Mastercycler nexus that are connected to each other must be in "Idle"
state. This also applies to all Mastercycler pro devices that are connected to a
control panel. No programs may be running.
Hint!
The current verification can be interrupted at any time using the Exit softkey.