For all other rotors and their rotor lids there is no service life limit as long as the following
conditions are met:
correct use
recommended care
undamaged condition
On rotors, the manufacturing date is embossed in the format 03/12 (= March 2012) and/or
the inside of the plastic lids, in the shape of a clock .
To ensure aerosol tightness, the following applies:
Exchange aerosol-tight rotor lids and caps after 50 autoclaving cycles.
For QuickLock rotor lids, replace the seal after 50 autoclaving cycles.
FA-45-48-11 5427 R/5430/5430 R
5804/5804 R/5810/
5810 R
100 000 7
FA-45-16-17 5430/5430 R 100 000 7
FA-45-12-17 5427 R 100 000 7
FA-45-20-17 5804/5804 R/5810/
5810 R
100 000 7
S-4-72 5804/5804 R/5810/
5810 R
100 000 7
S-4-104 5810/5810 R 100 000 7
S-24-11-AT 5427 R/5430/5430 R 100 000 7
T-60-11 5804/5804 R/5810/
5810 R
n/a 7
QuickLock rotor lid 3
Rotor lid and caps made
from polycarbonate (PC),
polypropylene (PP) or
polyetherimide (PEI)
––3
A-8-11 bucket 5417 R–3
Plastic adapters 1
Rotor/accessories Centrifuge Max. service life from initial start-up
in cycles in years