Register your instrument! www.eppendorf.
Copyright © 2013 Eppendorf AG, Hamburg. All rights reserved, including graphics and images. No part of this publication may be reproduced without the prior permission of the copyright owner. Trademarks Eppendorf® and the Eppendorf logo are registered trademarks of Eppendorf AG, Hamburg, Germany. Registered trademarks and protected trademarks are not marked in all cases with ® or ™ in this manual. 5821 900.
Table of contents Rolling Cabinet 3 English (EN) Table of contents Operating Rolling on see p. Tab. Fig. p. Cabinet manual English Operating (EN) manual 1 Operating instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2.
Table of contents 4 Rolling Cabinet English (EN) 9 Transport, storage and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.1 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.3 Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating instructions Rolling Cabinet 5 English (EN) 1 1.1 Operating instructions Using this manual Read the entire operating manual before using the rolling cabinet for the first time. Please also observe the centrifuge operating manual. This operating manual must be considered as essential part of the product and, thus, must always be easily accessible. If the rolling cabinet is transferred to a third party, this operating manual must be included with the cabinet.
Operating instructions 6 Rolling Cabinet English (EN)
Product description Rolling Cabinet 7 English (EN) 2 2.1 Product description Main illustration The figure shows the rolling cabinet version with integrated transformer. The centrifuge shown is not included in the delivery package. Abb. 2-1:Rolling cabinet with transformer 1 Fig.
Product description 8 Rolling Cabinet English (EN) 2.
Safety Rolling Cabinet 9 English (EN) 3 3.1 Safety Intended use The rolling cabinet is exclusively intended for Centrifuges 5810, 5810 R, 5804 and 5804 R and use indoors. The rolling cabinet is available in two versions: one version without electrical components and one version with electrical components (transformer). The rolling cabinet with transformer is exclusively intended for use with centrifuges (5810, 5810 R, 5804, 5804 R) with 120 V mains voltage.
Safety 10 Rolling Cabinet English (EN) 3.4 Warnings for intended use Read the operating manual and observe the following general safety precautions before using the rolling cabinet. As only one rolling cabinet version contains electronic components (transformers), the additional safety precautions for the rolling cabinet with transformer are listed separately. CAUTION! Poor safety due to incorrect accessories and spare parts.
Safety Rolling Cabinet 11 English (EN) Safety precautions only for rolling cabinet with transformer WARNING! Electric shock due to damage to device or mains cable. Only switch on the device if the device and mains cable are undamaged. Only use devices that have been properly installed or repaired. In case of danger, disconnect the device from the mains supply by pulling the power plug from the device or the mains socket or, by using the isolating device intended for this purpose (e.g.
Safety 12 Rolling Cabinet English (EN) NOTICE! Damage to plug connection. Pull the plug to disconnect the plug connection. Do not pull the cable. NOTICE! Radio interference. This device is a category A product in accordance with EN 55011. There may be disturbance to radio reception in residential areas. Ensure that appropriate preventive measures are taken. 3.
Installation Rolling Cabinet 13 English (EN) 4 4.1 Installation Selecting the location Select the rolling cabinet location according to the following criteria: • • • • Sturdy, level base at the location Distance to adjacent devices and walls: at least 30 cm Observe the location and positioning instructions in the centrifuge operating manual. Only rolling cabinet with transformer: mains power connection (100 V) according to name plate. The name plate is located on the rear side of the rolling cabinet.
Installation 14 Rolling Cabinet English (EN) 4.3 Installing the rolling cabinet Required tool and accessories • Allen key • Plastic plugs • Foot holders which match the centrifuge • Screws 8 foot holders are delivered with the rolling cabinet. • 4 foot holders, 48 mm in diameter for Centrifuge 5804 • 4 foot holders, 59 mm in diameter for Centrifuge 5810, 5810 R, 5804 R 4 foot holders must be bolted to the rolling cabinet to ensure the centrifuge is securely positioned on the rolling cabinet.
Installation Rolling Cabinet 15 English (EN) 4.3.1 Mounting the foot holders CAUTION! Risk of injury due to non-mounted foot holders or incorrect positioning of foot holders. The centrifuge may slip off the rolling cabinet if the centrifuge device feet are not in the foot holders. Use the foot holders that match the centrifuge type. Ensure that the boreholes match the centrifuge type. Center the centrifuge device feet in the foot holders. NOTICE! Damage due to penetrating liquid.
Installation 16 Rolling Cabinet English (EN) 6. Remove the adhesive film. 4.3.2 Installing the centrifuge on the rolling cabinet Prerequisites • The wheel stops are used to secure the castors and prevent the unit form rolling away. • The foot holders are mounted on the upper side of the rolling cabinet (see p. 15). CAUTION! Risk of crushing when installing the centrifuge on the rolling cabinet. When positioning the device feet into the foot holders there is a risk of crushing your fingers.
Installation Rolling Cabinet 17 English (EN) 4.3.3 Connecting the centrifuge (rolling cabinet with transformer) 1. Connect the centrifuge and the rolling cabinet with the connecting cable (see Main illustration on p. 7). The connecting cable is part of the centrifuge delivery package. 2. Connect the rolling cabinet to the mains power supply with the mains cable supplied. 3. Switch on the rolling cabinet using the mains power switch located on the right side of the rolling cabinet.
Installation 18 Rolling Cabinet English (EN)
Operation Rolling Cabinet 19 English (EN) 5 Operation CAUTION! Risk of injury due to non-secured castors Before starting a centrifugation, secure the castors with wheel stops. CAUTION! Risk of injury due to moving or bumping the device during centrifugation. Do not move or bump the rolling cabinet during centrifugation. Do not open the drawer during centrifugation. Only centrifuge when the castor stops are locked. CAUTION! Risk of tipping over when the drawer is extended.
Operation 20 Rolling Cabinet English (EN)
Maintenance Rolling Cabinet 21 English (EN) 6 6.1 Maintenance Cleaning and disinfection DANGER! Electric shock as a result of penetration of liquid. Switch off the device and disconnect the power plug before starting cleaning or disinfection work. Do not allow any liquids to penetrate the inside of the housing. Do not spray clean/spray disinfect the housing. Only plug the device back in if it is completely dry, both inside and outside. NOTICE! Damage from the use of aggressive chemicals.
Maintenance 22 Rolling Cabinet English (EN) The decontamination information must be observed when sending the device to the authorized technical service for repairs. 6.2 Decontamination before shipment If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer for disposal please note the following: WARNING! Risk to health from contaminated device 1. Follow the instructions in the decontamination certificate.
Troubleshooting Rolling Cabinet 23 English (EN) 7 7.1 Troubleshooting General errors If the suggested troubleshooting measures fail repeatedly, please contact your local Eppendorf partner. You can find the contact addresses on the Internet at www.eppendorf.com/worldwide. Only rolling cabinet with transformer: Symptom/ message Cause The centrifuge is • No mains power connection. not functioning properly. Remedy Check the rolling cabinet/ centrifuge connecting cable.
Troubleshooting 24 Rolling Cabinet English (EN)
Technical data Rolling Cabinet 25 English (EN) 8 8.1 Technical data Power supply Only rolling cabinet with transformer Power connection 100 V, 50 Hz – 60 Hz Current consumption Maximum 15 A Power consumption Maximum 1500 VA EMC: noise emission EN 61326 – class A EMC: noise immunity EN 61326 – class A Overvoltage category II Safety Overcurrent protective switch (18 A) 8.2 Weight/dimensions Dimensions Width 70.0 cm (27.6 in) Height 45.0 cm (17.7 in) Standard rolling cabinet Depth 60.
Technical data 26 Rolling Cabinet English (EN)
Transport, storage and disposal Rolling Cabinet 27 English (EN) 9 9.1 Transport, storage and disposal Transport The rolling cabinet may only be transported in its original packaging. Use a transport aid for to transport it long distances.
Transport, storage and disposal 28 Rolling Cabinet English (EN) In Germany, this is mandatory from March 23, 2006. From this date, the manufacturer has to offer a suitable method of return for all devices supplied after August 13, 2005. For all devices supplied before August 13, 2005, the last user is responsible for the correct disposal.
Ordering information Rolling Cabinet 29 English (EN) 10 Ordering information The rolling cabinet with transformer is only available in Japan. Order no. (International) 5821 020.028 5821 020.010 5804 000.722 5805 000.726 5810 000.726 5811 000.720 5821 851.027 5821 851.035 5821 851.043 5821 851.051 5821 851.060 Order no.
Ordering information 30 Rolling Cabinet English (EN) Order no. (International) 5821 851.078 5821 851.086 Order no.
Evaluate your manual Give us your feedback. www.eppendorf.com/manualfeedback Your local distributor: www.eppendorf.com/contact Eppendorf AG · 22331 Hamburg · Germany eppendorf@eppendorf.com · www.eppendorf.