Manual
6 Technical data
37
Eppendorf Research® plus — Operating manual
6.2 Research plus multi-channel variable
Research plus variable multi-channel
Model Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic Random
± % ± μL± % ± μL
0.5 – 10 L
Increment:
0.01 L
medium gray
0.1 - 20 L
40 mm
0.5 L ± 12 ± 0.06 ± 8.0 ± 0.04
1 L ± 8.0 ± 0.08 ± 5.0 ± 0.05
5 L ± 4.0 ± 0.2 ± 2.0 ± 0.1
10 L ± 2.0 ± 0.2 ± 1.0 ± 0.1
10 – 100 L
Increment:
0.1 L
yellow
2 - 200 L
53 mm
10 L ± 3.0 ± 0.3 ± 2.0 ± 0.2
50 L ± 1.0 ± 0.5 ± 0.8 ± 0.4
100 L ± 0.8 ± 0.8 ± 0.3 ± 0.3
30 – 300 L
Increment:
0.2 L
orange
20 - 300 L
55 mm
30 L ± 3.0 ± 0.9 ± 1.0 ± 0.3
150 L ± 1.0 ± 1.5 ± 0.5 ± 0.75
300 L ± 0.6 ± 1.8 ± 0.3 ± 0.9
6