Manual
5 Maintenance
33
Eppendorf Research® plus — Operating manual
Error limits in accordance with ISO 8655-2 - Research plus variable multi-channel
Model Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic Random
± % ± μL± % ± μL
0.5 - 10 L
Increment:
0.01 L
medium gray
0.1 - 20 L
40 mm
0.5 L ± 48.0 ± 0.24 ± 32.0 ± 0.16
1 L ± 24.0 ± 0.24 ± 16.0 ± 0.16
5 L ± 4.8 ± 0.24 ± 3.2 ± 0.16
10 L ± 2.4 ± 0.24 ± 1.6 ± 0.16
10 - 100 L
Increment:
0.01 L
yellow
2 - 200 L
53 mm
10 L ± 16.0 ± 1.6 ± 6.0 ± 0.6
50 L ± 3.2 ± 1.6 ± 1.2 ± 0.6
100 L ± 1.6 ± 1.6 ± 0.6 ± 0.6
30 - 300 L
Increment:
0.02 L
orange
20 - 300 L
55 mm
30 L ± 26.0 ± 7.8 ± 10.0 ± 3.0
150 L ± 5.2 ± 7.8 ± 2.0 ± 3.0
300 L ± 2.6 ± 7.8 ± 1.0 ± 3.0
5
Maintenance