Manual
5 Maintenance
32
Eppendorf Research® plus — Operating manual
30 - 300 L
Increment:
0.02 L
orange
20 - 300 L
55 mm
30 L ± 13.0 ± 3.9 ± 5.0 ± 1.5
150 L ± 3.0 ± 3.9 ± 1.0 ± 1.5
300 L ± 1.3 ± 3.9 ± 0.5 ± 1.5
100 - 1 000 L
Increment:
1 L
blue
50 - 1 000 L
71 mm
100 L ± 8.0 ± 8.0 ± 3.0 ± 3.0
500 L ± 2.0 ± 8.0 ± 0.6 ± 3.0
1 000 L ± 0.8 ± 8.0 ± 0.3 ± 3.0
0.5 - 5 mL
Increment:
5 L
purple
0.1 - 5 mL
120 mm
0.5 mL ± 8.0 ± 40.0 ± 3.0 ± 15.0
2.5 mL ± 1.6 ± 40.0 ± 0.6 ± 15.0
5.0 mL ± 0.8 ± 40.0 ± 0.3 ± 15.0
1 - 10 mL
Increment:
10 L
turquoise
1 – 10 ml
165 mm
1.0 mL ± 6.0 ± 60.0 ± 3.0 ± 30.0
5.0 mL ± 1.2 ± 60.0 ± 0.6 ± 30.0
10.0 mL ± 0.6 ± 60.0 ± 0.3 ± 30.0
Error limits in accordance with ISO 8655-2 - Research plus variable single-channel
Model Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic Random
± % ± μL± % ± μL
5
Maintenance