Manual
5 Maintenance
31
Eppendorf Research® plus — Operating manual
5.5.5 Error limits in accordance with ISO 8655-2
Error limits in accordance with ISO 8655-2 - Research plus variable single-channel
Model Test tip
epT.I.P.S. Color
code
Volume range
Length
Testing
volume
Error limits
Error
Systematic Random
± % ± μL± % ± μL
0.1 - 2.5 L
Increment:
0.002 L
dark gray
0.1 - 10 L
34 mm
0.1 L ± 100.0 ± 0.1 ± 75.0 ± 0.075
0.25 L ± 50.0 ± 0.125 ± 30.0 ± 0.075
1.25 L ± 10.0 ± 0.125 ± 6.0 ± 0.075
2.5 L ± 5.0 ± 0.125 ± 3.0 ± 0.075
0.5 - 10 L
Increment:
0.01 L
medium gray
0.1 - 20 L
40 mm
0.5 L ± 24.0 ± 0.12 ± 16.0 ± 0.08
1 L ± 12.0 ± 0.12 ± 8.0 ± 0.08
5 L ± 2.4 ± 0.12 ± 1.6 ± 0.08
10 L ± 1.2 ± 0.12 ± 0.8 ± 0.08
2 - 20 L
Increment:
0.02 L
light gray
0.5 - 20 L L
46 mm
2 L ± 10.0 ± 0.2 ± 5.0 ± 0.1
10 L ± 2.0 ± 0.2 ± 1.0 ± 0.1
20 L ± 1.0 ± 0.2 ± 0.5 ± 0.1
2 - 20 L
Increment:
0.02 L
yellow
2 - 200 L
53 mm
2 L ± 10.0 ± 0.2 ± 5.0 ± 0.1
10 L ± 2.0 ± 0.2 ± 1.0 ± 0.1
20 L ± 1.0 ± 0.2 ± 0.5 ± 0.1
10 - 100 L
Increment:
0.01 L
yellow
2 - 200 L
53 mm
10 L ± 8.0 ± 0.8 ± 3.0 ± 0.3
50 L ± 2.0 ± 0.8 ± 0.6 ± 0.3
100 L ± 0.8 ± 0.8 ± 0.3 ± 0.3
20 - 200 L
Increment:
0.02 L
yellow
2 - 200 L
53 mm
20 L ± 8.0 ± 1.6 ± 3.0 ± 0.6
100 L ± 2.0 ± 1.6 ± 0.6 ± 0.6
200 L ± 0.8 ± 1.6 ± 0.3 ± 0.6
5
Maintenance