Error Cause Solution
Droplets on the
inner wall of the
pipette tip.
Uneven wetting of the
plastic wall.
A pipette tip with poor
wetting properties has been
used.
Attach a new pipette tip.
Use an original Eppendorf
tip.
Pipette is
dripping and/or
the volume
pipetted is
incorrect.
The tip is loose.
A poorly fitting pipette tip
has been used.
Liquid with a high vapor
pressure has been pipetted.
For 5005000 µL/1 –10 mL
pipette:
Pipetting was too fast.
The tip was removed too
quickly from the liquid.
The pipette is dripping
because:
piston is contaminated.
Piston is damaged.
Seal is damaged.
The pipette lower part is
loose.
For multi-channel model:
The O-ring of the channel is
damaged.
Press the tip on firmly.
Use an original Eppendorf
tip.
In this case, we recommend
pipetting using a positive-
displacement pipette.
Allow the control button to
slide back slowly.
Pull the tip slowly out of the
liquid.
Clean and lightly lubricate
the piston (Part B,
"Maintenance").
Replace the piston and seal
(Part B, "Maintenance").
Replace the seal;
with the 5005000 µL and
1 10 mL model, replace
the piston (Part B,
"Maintenance").
Tighten the pipette lower
part by hand.
Replace defective O-ring
(Part B, "Maintenance").
7 Troubleshooting
Research – Part A – Section 7