User guide

83
Voltage/frequency: 100–240 V ±10 %, 50–60 Hz
Fuses: T0, 1.0 A – 5 x 20 mm (2 pcs.)
Power consumption: 60 W
Excess-voltage category:
II
Protection class:
I
Degree of contamination: 2
Voltage range
for the following modules:
Eukaryotic cells: 20– 100 V, in increments of 1 V
100–1,000 V, in increments of 10 V
1,000–1,200 V, in increments of 100 V
Interval between pulses: 1 minute
Cell fusion: 0, 5– 100 V, in increments of 1 V
100– 300 V, in increments of 10 V
Interval between pulses: 1 second
Alternating voltage: 0,
1.0– 10 V, in increments of 0.1 V
Bacteria: 200–1,000 V, in increments of 10 V
1,000–2,500 V, in increments of 100 V
No multiple pulse
Capacitor: 10 µF, 2,500 V impulse discharge
Resistance: 600
parallel
56
series
Charging time: <30 seconds
No. of pulses (n): 0–99
Time constant
τ
of discharge:
Eukaryotic cells: 15–500 µs, in increments of 5 µs,
fading e-function
Bacteria: Nominal 5 ms,
with an impedance of the sample of 3.3 k
Measurement and display from 0.8–6.0 ms
Duration of alignment: 0–95 s, in increments of 5 s
Duration t of pulse:
Cell fusion: 0, 15 µs–300 µs, in increments of 5 µs,
square-wave pulse
RS 232 interface: 9,600 baud, 8 bits, no parity,
2 stop bits, XON/XOFF
Ambient temperature: max. 40 °C
Relative ambient humidity: max. 80 %
8 Technical data