Copyright Copyright© 2012 New Brunswick Scientific Co., Inc., USA. No part of this publication may be reproduced without the prior permission of the copyright owner. New Brunswick reserves the right to change information in this document without notice. Updates to information in this document reflect our commitment to continuing product development and improvement. Trademarks Eppendorf® is a registered trademark of Eppendorf AG, Germany.
Galaxy O2 Sensor Setup — Instructions 1 1 1 Galaxy O2 sensor setup Galaxy O2 sensor setup 1.1 Galaxy O2 sensor setup The O2 sensor used on all New Brunswick Galaxy incubators equipped with the O2 option is a sensitive, self-powered, electrochemical cell, with a typical life span of one to two years dependent on ambient atmospheric levels. To ensure the maximum shelf life of the O2 sensor, this incubator is equipped with a placeholder O2 sensor.
Galaxy O2 Sensor Setup — Instructions 1 1.2 Removing existing O2 sensor Galaxy O2 sensor setup 1. Remove the rear access cover plug to gain access to the O2 sensor, (On older models, the O2 sensor can be accessed by removing the six M4 x 12 screws or the four M4 x 12 screws) 2. Unplug the connector to disconnect the sensor. Hint! 4 Avoid pulling the wire leads. Hold the white connector body when disconnecting the sensor.
Galaxy O2 Sensor Setup — Instructions 3. 1 Insert sensor tool and turn counter-clockwise to remove O2 sensor. Galaxy O2 sensor setup O2 sensor removal completed. 1.3 Installing O2 sensor on Galaxy 170 liter or 48 liter incubators 1.
Galaxy O2 Sensor Setup — Instructions 1 Galaxy O2 sensor setup 2. Insert the O2 sensor into the access port using the O2 sensor tool. Hint! Push O2 sensor into access port; ensuring the connector and wire have been taken out and held to the side. 3. Turn clockwise to secure. Hint! 6 Finger tighten the O2 sensor only, do not use metal tools or excessive force.
Galaxy O2 Sensor Setup — Instructions 1 4. Connect the sensor by plugging in the connector. Galaxy O2 sensor setup Hint! Hold the white plug and NOT the wires; connect the Incubator to the O2 sensor and ensure that the connector is fully inserted and oriented correctly 5. Replace the rear access cover. Ensure that access port cover plug is fully pushed into place. Hint! 6. Allow the humidified incubator to stabilize overnight.
Galaxy O2 Sensor Setup — Instructions 1 1.4 Galaxy O2 sensor setup Hint! Installing O2 sensor on Galaxy 14 liter incubator • Before replacing the O2 sensor disable the O2 control as follows: a b c Press the programming ( * ) key. Select ‘02C.0nt’ using the » and º keys and press the ENTER key. Toggle to the OFF position and press the ENTER key. 1. Remove the O2 sensor access plug at the rear of the incubator. Twist and Pull access cover plug to remove. Hint! 2.
Galaxy O2 Sensor Setup — Instructions 1 Hint! Push O2 sensor into access port; ensuring the connector and wire have been taken out and held to the side Hint! Galaxy O2 sensor setup 3. Turn clockwise to secure. Finger tighten the O2 sensor only, do not use metal tools or excessive force. 4. Connect the sensor by plugging in the connector.
Galaxy O2 Sensor Setup — Instructions 1 Galaxy O2 sensor setup 5. Replace the rear access cover. Ensure that access port cover plug is fully pushed into place. Hint! 6. Allow the humidified incubator to stabilize overnight. 1.5 Referencing the O2 sensor for Galaxy 170 liter and 48 liter incubators Reference the O2 sensor to atmospheric O2 levels as follows: 1. Select ‘O2 sensor - REF TO ATMOSPHERE’ which can be located in the USER Menu. 2. Follow the on-screen instructions.
Galaxy O2 Sensor Setup — Instructions 1 Galaxy O2 sensor setup 60 second countdown begins when user opens door.
Galaxy O2 Sensor Setup — Instructions 1 10 minute countdown begins when user closes door. Galaxy O2 sensor setup Hint! The oxygen reading is automatically adjusted to 19.7 % which is the true reading taking into account the Relative Humidity level. The incubator is now ready for use with O2 control. Hint! 1.6 Additional calibration is required for 170 liter and 48 liter incubators with the 0.
Galaxy O2 Sensor Setup — Instructions 1.7 1 Calibrating the O2 sensor for Galaxy 170 liter and 48 liter 0.1 - 19 % O2 option Galaxy O2 sensor setup Requirement • Nitrogen source capable of supplying clean gas at 5 PSI (0.35 Bar) • 6 mm bore PVC tubing The following instructions applies to all incubators equipped with the 0.1 - 19 % O2 option. 1. Re-reference the O2 sensor (see Referencing the O2 sensor for Galaxy 170 liter and 48 liter incubators on p. 10). 2. Program the incubator for 0 % O2. 3.
Galaxy O2 Sensor Setup — Instructions 1 Abb. 1:Connection schematic 2 3 Galaxy O2 sensor setup 1 Fig.
Eppendorf offices
Your local distributor for New Brunswick products: www.nbsc.com/ContactUs New Brunswick Scientific · 175 Freshwater Blvd., Enfield, CT 06082-4444 · USA Eppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 01-556 · E-mail: eppendorf@eppendorf.com New Brunswick Scientific Europe B.V. · Nijmegen · The Netherlands · Tel: +31 (0) 24 3717 600 · Email: europe@nbsbv.nl Eppendorf North America, Inc. · Hauppauge, N.Y. USA · Tel: +1 516 334 7500 · +1 800 645 3050 · E-mail: info@eppendorf.