8.2.6 Viewing the properties of the labware
Proceed as follows:
Prerequisites
The Labware Editor is open.
Select the labware.
The properties of the labware are displayed in the Properties column.
8.2.7 Setting the bottom tolerance
Proceed as follows:
Prerequisites
The Labware Editor is open.
1. Select the labware.
2. Click on the symbol.
The Edit Properties window appears.
Abb. 8-6: Edit Properties window
Fig. 8-6: Edit Properties window
NOTICE! Material damage due to placing the pipette tip on the base of the vessel.
If the pipette tips touch the bottom of the tube, the bottom tolerance is too low. The labware
will become damaged and the dispensing volume will be distorted.
Test the application without liquid.
Check the bottom tolerance when exchanging pipette tips, plates and vessels.
If pipette tips become jammed at the base of the vessel, increase the bottom tolerance.
If deviations in the dispensing volume occur, increase the bottom tolerance.