6.5.1.1 Setting application parameters for the Detect volumes function
Abb. 6-34: Properties window
Fig. 6-34: Properties window
Tab. 6-5: Detect volumes
To set the application parameters for the Detect volumes function (see Defining the application
parameters of the labware on p. 44).
Off • Deactivate the optical sensor.
• The volume of all vessels must be entered when the application is
started.
For labware with ≤ vessels, the volume for each vessel can be
specified in the Volume column.
For labware with > 56 vessels, one volume can be specified for all
vessels. This volume will be recommended at the start of the
application, and can be changed.
• The volume specification does not take into account the surface
meniscus of the liquid.
• If the volume is not enough, the result may be faulty dispensing.
All positions • Check the volume of all positions in a labware.
• If you work with an eight-channel dispensing tool, the epMotion uses
the calculated volume for each vessel.
• If you work with a multi-channel dispensing tool, the epMotion uses
the lowest calculated volume in a column for the liquid aspiration
and the highest calculated volume in a column for the liquid
dispensing.
If the filling levels in one plate differ significantly or the labware has very low filling levels, use
the Aspirate from bottom and Dispense from top option. Do not use the Random access (see
Sample transfer > Options command on p. 50) option.