Owner manual
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
74
6.3.17.2 Importing a CSV file
Prerequisites
• CSV file available (see p. 72).
Import a CSV file as follows:
1. Create a new application.
2. Insert the Number of samples command as the first command.
3. In the Number of samples command, activate the Fixed sample count checkbox.
4. Set the Number of samples parameter to 1.
5. Insert the Sample transfer command as the second command.
6. In the command Sample transfer >, determine parameter locations for source labware and destination
labware.
The locations defined in the Sample transfer command are available for the commands in the CSV file.
4 locations each are available for the source labware and the destination labware.
In the example, 2 locations each were defined for the source labware and destination labware in the
Sample transfer command. These locations are available for commands in the CSV file.
The number of the labware from the CSV file must correspond to the parameters for the Sample transfer
command.
Labware is used in the sequence defined in the Sample transfer command. If rack 2 is specified in the
CSV file in column B, the second rack from the Source area will be used as the source vessel for this
command.
A maximum of 500 transfer commands can be imported.
The settings for all commands in the CSV file are defined in the Sample transfer command.