Owner manual
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
69
6.3.13.1 Transporting the Vac Lid with the gripper
Use the Vacuum command to transport the Vac Lid from location T0 to the filter plate in the vacuum unit
and back to location T0 with the gripper.
Prerequisites
• The Vac Lid is in location T0.
To have the Vac Lid transported from location T0 to the filter plate at the start of the Vacuum command,
activate the Use Vacuum Lid checkbox.
After the time set in the Duration field has elapsed, the gripper takes the Vac Lid off the filter plate at an
underpressure of 100 mbar (10 kPa). The gripper transports the Vac Lid to position T0.
6.3.14 Transport command
The Transport command can be used to transport labware. The gripper takes labware from one location and
transports it to another location.
Abb. 6-23: Transport window
Fig. 6-23: Transport window
The command Transport only is available for devices with gripper.
Labware
Select the labware.
To location
Select a location.
Information on the labware that is being transported can be found in the hardware operating
manual.