6.3.13 Vacuum command
The Vacuum command is used to create a vacuum in the vacuum unit. You can aspirate liquid in filter
plates.
Prior to executing the application, equip the vacuum unit as described in the epMotion hardware operating
manual.
Abb. 6-22: Vacuum window
Fig. 6-22: Vacuum window
The command Vacuum only is available for devices with vacuum unit.
Frame drop-down list
Type of Vac Frame used.
Pressure field
Vacuum in the vacuum unit. Unit mbar or kPa.
Maximum underpressure 850 mbar/85 kPa.
Duration field
Time during which the vacuum is maintained.
0 s – 100 min 59 s range.
Use Vacuum Lid checkbox
Activate transport of the Vac Lid with the gripper.
Use Vacuum Lid drop-down list
Vac Lid used.
Check levels checkbox
The optical sensor checks the remaining volume
of all wells in a filter plate on the vacuum unit. If
remaining liquid has been detected, a vacuum
will be reapplied. Checking the remaining volume
takes too long.
BAR