Owner manual

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
67
Tab. 6-2: Maximum speed for labware on the thermomixer
6.3.12.1 Setting the template for the Thermomixer
Prerequisites
The Thermomixer command is selected (see p. 66).
1. Edit the template in the Options field.
The word Changed appears next to the Save button.
2. Enter the name of the template.
3. Click on the Save button.
The template will be saved.
6.3.12.2 Managing templates for Thermomixer
Abb. 6-21: Labware Editor > TMX Templates
Fig. 6-21: Labware Editor > TMX Templates
Manage the templates for the thermomixer in the Labware Editor > TMX Templates.
Labware Maximum speed in rpm
Tubes 1000
DWP 96 1200
DWP 384 1200
MTP 6 1000
MTP 96 2000
MTP 384 2000
PCR 96 2000
PCR 384 2000
Reservoir Rack Cannot be mixed
Thermorack/Thermoblock 1000
Thermoadapter 96 1000
Thermoadapter 384 1000
Thermorack TMX 1300
PrepRack 1300