epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
66
6.3.12 Thermomixer command
The Thermomixer command can be used to set the temperature and speed of the thermomixer.
Abb. 6-20: Thermomixer window
Fig. 6-20: Thermomixer window
The command Thermomixer only is available for devices with thermomixer.
Template drop-down list
The templates for the labware contain saved
settings for mixing.
Mixing on checkbox
Activate the mixing function of the thermomixer.
Speed field
Set the mixing frequency. 300 rpm 2000 rpm
range.
Duration field
Set the mixing time. 5 s 120 min range.
Temperature on checkbox
Activate the temperature control of the
thermomixer.
Keep temperature after applicaton run field
Activate the temperature control for the period
after the end of the application. This function can
be used to cool samples or reagents after the end
of the application.
Temperature field
Set the set temperature of the thermomixer,
4 °C 95 °C range.
Information on the thermomixer can be found in the epMotion hardware operating manual.
NOTICE! Contamination of samples and the device due to high mixing frequency.
If the speed is set too high, liquid will spray out of the vessels. Labware may become loose
from the holder and fly about.
Observe the maximum speeds of the labware in the operating instructions.