Owner manual

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
63
If the Magnetic separation command is executed when the Separation On checkbox is checked, the
magnetic fingers will be drawn out. During the waiting period, the magnetic beads will be separated by the
drawn out magnetic fingers on the tube inner wall. After the waiting period has elapsed, the next command
in the procedure will be executed. If the Magnetic separation command is executed when the Separation On
checkbox is not checked, the magnetic fingers will be drawn in.
Abb. 6-17: Magnetic separation command
Fig. 6-17: Magnetic separation command
Starting the separation
To start the separation, execute the Magnetic separation command when the Separation On checkbox is
activated.
The magnetic fingers will be drawn out.
Exiting the separation
To end the separation, execute the Magnetic separation command when the Separation On checkbox is
deactivated.
The magnetic fingers will be drawn in.
6.3.9 Mix command
The Mix command can be used to mix liquids in a position by multiple aspiration and dispensing steps.
The options of the Mix command are identical with the options of the Sample transfer command in the Mix
area (see p. 51).
Separation enabled checkbox
Check the checkbox to start the separation.
Uncheck the checkbox to end the separation.
Duration field
Waiting period. The application stops until the
waiting period has elapsed.