Owner manual

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
58
6.3.3.1 Example of a dilution series
This example explains the concept of a dilution series. The example does not describe the full application.
Objective
24 samples are in a rack with 24 positions and should be diluted 1:1000.
The dilution should be completed in 3 program steps with dilutions of 1:10.
First, a single-channel dispensing tool is used. An eight-channel dispensing tool will then be used to
reach a higher speed.
Procedure
200 µL of each of the samples will be transferred to the 96-well plate using the Sample transfer
command. The dilutions are carried out in this plate.
225 µL of the diluent will be transferred from the 300 mL reservoir to each of the empty wells of the
96-well plate using the Reagent transfer command.
•25 µL of the sample will be aspirated from column 1 using the Dilute command and will be mixed with
respectively 225 µL of the diluent in the columns 2, 3 and 4. The mixing process will be repeated twice
in the following columns.
In this example, each dilution step is a 1:10 dilution. The required 1:1000 dilution will be achieved with the
third 1:10 dilution. The volume aspirated from the undiluted sample also applies for the dilution steps.
Creating dilution series
Transferring undiluted samples to the destination labware
1. Pattern of the Sample transfer command:
The source position and destination position are the same color. epBlue detects the pattern and
suggests the next field.