Owner manual

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
53
Protein Water with
1 % albumin
(10 g/L),
0.01 % Triton
X-100
5 µL 5 µL New pipette tips are
pre-wetted with the liquid to
be aspirated.
Prewetting option is set (see
p. 54).
The bending of the surface of
the liquid has a negative
impact on the free jet
capability of dispensing in cell
culture plates.
Protein C Water with
1 % albumin
(10 g/L),
0.01 % Triton
X-100
5 µL 5 µL New pipette tips are
pre-wetted with the liquid to
be aspirated.
Prewetting option is set (see
p. 54).
Dispensing the liquid with a
greater distance to the
calculated plane surface of the
liquid (4 to 5 mm) than liquid
type Protein.
All other data such as liquid
type Protein.
Recommended for nutrient
media.
Rinse Demineralized
water.
Water with low
surfactant
content.
1 µL 3 µL Low aspiration and dispensing
speed, strongly delayed
blow-out.
Other data such as liquid type
Water.
This liquid type is suitable for
the mixing function of a
transfer command or if the Mix
command is used.
Can increase the risk of
contamination for small
vessels, e.g., PCR plates.
Recommended in combination
with the Mix option for
reducing the residual moisture
in the pipette tip.
Liquid type Optimized for 50 µL tips:
pipetting as of
50 µL tips:
dispensing as of
Remarks