Owner manual
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
53
Protein Water with
1 % albumin
(10 g/L),
0.01 % Triton
X-100
5 µL 5 µL • New pipette tips are
pre-wetted with the liquid to
be aspirated.
Prewetting option is set (see
p. 54).
• The bending of the surface of
the liquid has a negative
impact on the free jet
capability of dispensing in cell
culture plates.
Protein C Water with
1 % albumin
(10 g/L),
0.01 % Triton
X-100
5 µL 5 µL • New pipette tips are
pre-wetted with the liquid to
be aspirated.
Prewetting option is set (see
p. 54).
• Dispensing the liquid with a
greater distance to the
calculated plane surface of the
liquid (4 to 5 mm) than liquid
type Protein.
• All other data such as liquid
type Protein.
• Recommended for nutrient
media.
Rinse Demineralized
water.
Water with low
surfactant
content.
1 µL 3 µL • Low aspiration and dispensing
speed, strongly delayed
blow-out.
• Other data such as liquid type
Water.
• This liquid type is suitable for
the mixing function of a
transfer command or if the Mix
command is used.
• Can increase the risk of
contamination for small
vessels, e.g., PCR plates.
• Recommended in combination
with the Mix option for
reducing the residual moisture
in the pipette tip.
Liquid type Optimized for 50 µL tips:
pipetting as of
50 µL tips:
dispensing as of
Remarks