Owner manual

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
50
6.3.1.2 Sample transfer > Options command
Abb. 6-9: Sample transfer > Options
Fig. 6-9: Sample transfer > Options
Aspirate from bottom Aspirate liquid from the base of the vessel. When using several vessels with
different volumes, enter the lowest filling volume of these vessels (Sample
transfer > Parameter).
Elution from filter Aspirate liquids from the supernatant in filter plates. Available if in the
Parameter window the Pipette radio button is selected.
Dispense from top During liquid dispensing, position the pipette tips 3 mm – 4 mm below the
edge of the vessel. When using several vessels with different volumes, enter
the highest filling volume of these vessels (Sample transfer > Parameter).
Change Tips Defined time when the pipette tips are changed.
when command finished Eject pipette tips after the end of the command. The liquid will be aspirated
with the same tip from all source labware positions.
before asp. for next
destination well
Eject the pipette tips before the liquid from the next source position is
aspirated. Liquid from the same source position will be aspirated with the same
tip. Liquid from the next source position will be aspirated using a new tip.
before each aspiration Eject pipette tips prior to each aspiration. Liquid will always be aspirated with
a new tip. Additional volume will be discarded in the waste container along
with the pipette tip.
keep tips, do not change
tips
Do not eject pipette tips. Liquid will be aspirated with the same tip from all
source positions. The tips are also used in the next command. If the transfer
type changes in one of the following commands, or the Mix command is used,
the tips will be ejected.
after XX aspirations Pipette tips will be ejected after a defined number of aspirations.
Restore tips Pipette tips will be placed back into the position of the box they were removed
from.
Is only executed if the tips within an application are reused for the same source
vessel.
The restore tips checkbox is only available with specific settings of the Change
Tip option.