Owner manual
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
49
6.3.1.1 Sample transfer > Parameter command
Abb. 6-8: Sample transfer > Parameter
Fig. 6-8: Sample transfer > Parameter
Volume Select volume in µL or nL. A decimal place is available for volumes ≤ 99.9 µL.
You can transfer a volume larger than the maximum volume of the pipette tip.
The volume to be transferred will be pipetted in several steps.
Example: The user would like to transfer 1500 µL and enters this value into the
Volume field. If a 1000 µL tip is being used, 2x 750 µL will be pipetted. If a
300 µL tip is being used, 5 x 300 µL will be pipetted.
Pipet. tool Used dispensing tool.
Pipette The volume is individually aspirated and dispensed for each liquid transfer.
Multidispense Aspirate liquid and dispense it in the destination position in defined subsets.
Multiaspirate Aspirate liquid from several source positions using one pipette tip and
dispense the liquid in one destination position. Available for Pool and Pool one
destination commands.
Use filter tips Activate when using tips with filter.
Source Select the source labware. The source labware must be placed on the
worktable. You can select 4 source labwares. If you are using more than one
source labware, the number of positions in the labware (384, 96, 24) must
match. If you dispense within one labware, the source labware and destination
labware are identical.
Destination Select the destination labware. The destination must be placed on the
worktable. You can select 4 destination labwares. If you are using more than
one destination labware, the number of positions in the labware (384, 96, 24)
must match. If you dispense within one labware, the source labware and
destination labware are identical.
Pattern Select the source positions and destination positions. Create replicates.
Information on the aspiration and transfer of liquids by means of the epMotion can be found in
the hardware operating manual.