On the vacuum unit, you can stack 4 labware components. Information on the vacuum unit can be found in
the hardware operating manual.
Abb. 6-4: Stacking labware on the epBlue worktable
Fig. 6-4: Stacking labware on the epBlue worktable
Proceed as follows to stack labware at a location.
1. Place lower labware at one location.
2. Place upper labware at the same location.
The Placed labware column shows, which labware is placed in the epBlue worktable locations.
The number behind the designation shows how many parts have been stacked at this location. Example:
pcr384_3. Only 3 plates have been stacked.
Plates that are stacked may only be filled up to the working volume.