Owner manual

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
42
Prerequisites
Labware combinations have been created for reaction vessels (see p. 92).
The application is open (see p. 25).
The Worktable window is open.
1. Select labware from the Labware area.
2. Draw the labware to the desired position of the epBlue worktable.
If the labware cannot be positioned at this location, the location is marked in red.
6.2.5.2 Deleting positioned labware from the epBlue worktable
There are several options to delete labware from the epBlue worktable.
Selecting the labware
Activate the checkbox for the corresponding labware in the Placed labware column.
Alternatively, select the labware with the mouse.
Deleting the labware
Drag the selected labware into the trash with the mouse.
Alternatively, click on the symbol in the menu bar.
Or click on the selected labware with the right mouse button and select the Delete function in the
context menu.
6.2.5.3 Stacking labware
The gripper stacks and transports labware. The gripper is required to stack labware or to unstack a stack
during an ongoing application.
You can stack a maximum of 5 plates on the worktable. You can stack a plate on a height adapter. The
maximum stacking height is 126 mm. Stacked labware must have the same properties, e.g. name,
geometry, and bottom tolerance.
The gripper only transports one part at a time. The gripper can stack plates or unstack a stack of plates.
You can stack the following plates:
EP_pDNA_96_MTP
EP_TT_PCR_150
EP_TT_PCR_40
EP_DWP_1200
EP_pDNA_96_DWP
EP_DWP_200
Liquids cannot be transferred in a plate stack.
Functions of the optical sensor in case of stacked plates
The Check labware placement function is available.
The Detect volumes function is not available.