0x2135 Location for
destination not
defined
Edit mode/parameter in
command Transport:
The destination location for
the labware to be transported
was not edited.
See explanation in
"Cause".
0x2136 Invalid entry for
seconds (1 ... 59)
Edit mode/parameter in
command Wait:
A value beyond the allowed
range has been entered for the
seconds parameter.
Enter a value between
0 and 59 seconds.
0x2137 Invalid entry for
minutes (1 ... 99)
Edit mode/parameter in
command Wait:
A value above the allowable
maximum minutes has been
entered for the minutes
parameter.
Enter a value between
0 and 99 minutes.
0x2138 Method without
active commands
Application only contains
passive commands (such as
wait, comment, etc.).
Insert at least one active
command.
0x2139 Parameter conflict:
mix before
aspirating not
allowed in
multiaspirate mode
Pool/Pool One Destination:
Mix before aspirating not
allowed in multiaspirate mode.
Do not mix.
0x2152 Editing vac lid
option in vacuum
command not
possible because
no vac lid on the
worktable
To use the Vac Lid option in
the vacuum command the
worktable must be equipped
with a vac lid.
Editing the worktable
please place a Vac Lid on
position "GRIPPER".
0x2154 Parameter conflict:
Check levels time
may not be shorter
than remove lid
time
The check levels option can
only be performed when the
lid has been taken off the filter
plate before by the gripper.
Select a check levels time
greater than the remove lid
time.
0x2170 Parameter conflict:
Parameter elution
from filter is only
possible when filter
plates have been
selected as source
To edit the elution from filter
option a filter plate must have
been edited as source.
See explanations in
"Cause".