epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
171
0x2127 Mixing volume in
destination not
defined
Parameter Volume in a mix
procedure for destination
vessels was not edited in the
application (mix procedure as
part of a liquid transfer
command via parameter
Options).
See explanation in
"Cause".
0x2128 Parameter conflict:
mixing volume in
destination too
large for this tool
The Volume and Pipet. Tool
parameters in a mix procedure
for destination vessels do not
match (Volume is higher than
the upper limit of the tool
volume range) (mix procedure
as part of a liquid transfer
command via parameter
Options).
Edit Volume and Pipet. Tool
so that Volume is within
the pipette tool's volume
range.
0x2129 Parameter conflict:
mixing volume in
destination too
small for this tool
The Volume and Pipet. Tool in
a mix procedure for
destination vessels do not
match (Volume is lower than
the minimum allowed volume)
(mix procedure as part of a
liquid transfer command via
parameter Options).
Edit Volume and Pipet. Tool
so that Volume is within
the volume range of the
pipet tool.
0x212A Parameter conflict:
mixing volume in
destination greater
than filling volume
of destination tube
or well
The Volume and Destination
parameters in a mix procedure
for destination vessels do not
match (Volume is higher than
the maximum filling volume of
the destination vessel) (mix
procedure as part of a liquid
transfer command via
parameter Options).
Edit Volume and
Destination so that Volume
is within the maximum
filling volume of the
destination vessel.
0x212C Parameter conflict:
mix after dispense
not allowed in
multidispense
mode
When the transfer type
parameter is set to
multidispense, the mix after
dispense parameter cannot be
edited for this command.
Change parameter transfer
type to pipette, or:
Omit the mixing step; in
this case you could also
edit another mixing step as
a new command (Mix),
which would be performed
after the previous
command of the procedure
has ended.
Code Symptom/message Cause Remedy